Technical specifications
General |
|
Dimensions |
ca. 29 x 8 x 22 cm (W x L x H) |
Weight |
ca. 2000 g |
Operational voltage power unit |
100-240V~ / 50-60 Hz |
Power consumption, typ. |
30 W |
MDD 93/42/EEC classification |
IIa |
Protection against electric shock |
Protection class II with functional earth |
Protection level of applied part |
Type BF |
Ingress protection |
IPX0 |
Operating temperature |
+10 to +35 °C |
Storage temperature |
-20 to +70 °C |
Operating / Storage humidity |
20 to 85% (non-condensing) |
Operating height |
0 - 2000 m |
Type of operation |
Continuous operation |
Inflation time (0 to 180 mmHg/Operation at |
8 s |
Operating threshold for safety valve |
320 ± 10 mmHg |
Max. leakage rate |
6 mmHg / min |
Standards used |
EN 60601-1, EN 1060-1, EN 1060-3 |
Blood pressure |
|
Measurement range - systole |
30 to 280 mmHg |
Measurement range - diastole |
10 to 240 mmHg |
Measurement range - MAP |
20 to 255 mmHg |
Measurement range - pulse |
30 to 240 1/min |
Pressure measurement range |
10 to 330 mmHg |
Pressure resolution |
1 mmHg |
Pulse resolution |
1/min |
Precision of cuff pressure (static) |
± 3 mmHg |
Precision of blood pressure measurement |
< ± 5 / 8 mmHg (Mean / Standard deviation) |
Precision of pulse measurement |
2/min or ± 2% of displayed value (the larger of the |
Duration of a measurement cycle without |
26 s |
Ankle-Brachial Index (ABI) |
|
Measurement range - ABI |
0,10 to 9,33 |
ABI resolution |
0,01 |
Brachial-ankle pulse wave velocity (baPWV) |
|
Measurement range - baPWV |
6 to 999 m/s |
baPWV resolution |
0,1 m/s |
Resolution of baPWV curves on time scale |
5 ms |
Interfaces |
|
LAN |
RJ45, Ethernet 10/100 Mbit/s, reinforced insulation |
USB |
currently not used (later option) |
Service |
currently not used (later option) |