Technical specifications

 

 

General

Dimensions

ca. 29 x 8 x 22 cm (W x L x H)

Weight

ca. 2000 g

Operational voltage power unit
GS-1482
Operational voltage VascAssist

100-240V~ / 50-60 Hz
24V= / 2.1A with functional earth

Power consumption, typ.

30 W

MDD 93/42/EEC classification

IIa

Protection against electric shock

Protection class II with functional earth

Protection level of applied part

Type BF

Ingress protection

IPX0

Operating temperature

+10 to +35 °C

Storage temperature

-20 to +70 °C

Operating / Storage humidity

20 to 85% (non-condensing)

Operating height

0 - 2000 m

Type of operation

Continuous operation

Inflation time (0 to 180 mmHg/Operation at
sea level/closely fitting standard blood
pressure cuff) typ.

8 s

Operating threshold for safety valve

320 ± 10 mmHg

Max. leakage rate

6 mmHg / min

Standards used

EN 60601-1, EN 1060-1, EN 1060-3

 

 

 

Blood pressure

Measurement range - systole

30 to 280 mmHg

Measurement range - diastole

10 to 240 mmHg

Measurement range - MAP

20 to 255 mmHg

Measurement range - pulse

30 to 240 1/min

Pressure measurement range

10 to 330 mmHg

Pressure resolution

1 mmHg

Pulse resolution

1/min

Precision of cuff pressure (static)

± 3 mmHg

Precision of blood pressure measurement
(dynamic)

< ± 5 / 8 mmHg (Mean / Standard deviation)

Precision of pulse measurement

2/min or ± 2% of displayed value (the larger of the
values)

Duration of a measurement cycle without
artifacts (without inflation time / RR 120 :
80 / pulse 80), typ.

26 s

Ankle-Brachial Index (ABI)

Measurement range - ABI

0,10 to 9,33

ABI resolution

0,01

Brachial-ankle pulse wave velocity (baPWV)

Measurement range - baPWV

6 to 999 m/s

baPWV resolution

0,1 m/s

Resolution of baPWV curves on time scale

5 ms

Interfaces

LAN

RJ45, Ethernet 10/100 Mbit/s, reinforced insulation

USB

currently not used (later option)

Service

currently not used (later option)